| TYPE | Insulation resistance meter | 
                  
                    | PHYSICAL SPECIFICATIONS | 19”W x 7”H x 15”D, (48 cm x 17 cm x 38 cm); Weight: 24lbs (10.9 kg) | 
                  
                    | INPUT POWER | 110 – 120 Vac or 220 – 240 Vac (selectable), 50/60Hz | 
                  
                    | RESISTANCE READING RANGE (0°C to +30°C)
 | 100 K-ohms – 1 M-ohm ±20%, 1 M-ohm – 1 T-ohm ±5%,
 1 T-ohm – 5 T-ohm ±20% , 5kV
 100 K-ohms – 1 M-ohm ±20%,
 1 M-ohm – 100 G-ohms ±5%,
 100 G-ohms – 500 G-ohms ±5%, 500 V
 1 M-ohm – 10 G-ohms ±5%, 50 V
 | 
                  
                    | TEST VOLTAGE | Selectable from 50Vdc – 5 KVdc, in 2 Vdc steps | 
                  
                    | OUTPUT VOLTAGE ACCURACY (0°C to +30°C)
 | ±2%, ±1v of selected voltage with load resistance greater than 100 MΩ | 
                  
                    | SHORT CIRCUIT CURRENT | 2 mA max | 
                  
                    | CURRENT READING RANGE (0°C to +30°C)
 | 0.03 nA – 2 mA; Accuracy: ±5% ±0.2 nA | 
                  
                    | CAPACITANCE READING RANGE (0°C to +30°C)
 | 0.01 μF – 10.0 μF (Test voltage greater than 100V); Accuracy: ±5% ±0.03 μF | 
                  
                    | VOLTAGE READING RANGE (0°C to +30°C)
 | 50 – 1250 Vac (rms) or dc; Accuracy: ±5%, ±1V | 
                  
                    | CAPACITOR DISCHARGE | Less than 2 Sec/μF, automatic at the end of test | 
                  
                    | HUM REJECTION | 1mA per 1kv of test voltage, 2mA rms maximum | 
                  
                    | BATTERIES | Two 12V, 2.0 Ah sealed lead acid batteries. Battery life: typical 6 hrs, continuous testing. The IRM-5000P can be used while charging. | 
                  
                    | PRINTER | Built-in 2.5-inch wide thermal printer can print test results in both tabulated and graphic formats | 
                  
                    | COMPUTER INTERFACE | One RS-232C port | 
                  
                    | SAFETY |  | 
                  
                    | ENVIRONMENT | Operating: -10 to 50 C (15 to +122 F); Storage: -30 C to 70 C (-22 to +158 F) | 
                  
                    | CABLES | One 6-foot cable set, one 15-foot cable set, ground cable, power cord, cable bag | 
                  
                    | OPTIONS | Transportation Case | 
                  
                    | WARRANTY | One year on parts and labor | 
                  
                    | Specifications are valid at nominal voltage and ambient temperature of +25C (+77F). Last updated on February 6, 2009
 |